Last night Operations performed some rigorous tests on the effect of various
magnets on the target holding field. All of these magnet systems steer the
electron beam. Below is a summary of these results.
_____________________________________________________________________________________________
STATE DEVICE MEASUREMENT NOTES
BLAST HOLDING BLAST BRP TRANS LONG
MAGNET FIELD TRIMS MAGNET FIELD FIELD
_____________________________________________________________________________________________
ON 10.1 -3.7 Residual
magnetization in iron
OFF X X X X or field probe
calibration offsets
_____________________________________________________________________________________________
ON X 6.9 -3.4 small effect
OFF X X X
_____________________________________________________________________________________________
ON X 216.8 250.8 Should be
250.0 trans
OFF X X X
_____________________________________________________________________________________________
ON X 10.2 -3.6 No change due
to BRP
OFF X X X when BLAST
field is OFF
_____________________________________________________________________________________________
ON X 28.4 -4.9 18.3 gauss to
trans
OFF X X X from BLAST
magnet
_____________________________________________________________________________________________
ON X X 29.5 -4.9 1.3 gauss
change in trans
OFF X X when BLAST is
ON
_____________________________________________________________________________________________
ON X X X 28.2 -4.8
OFF X
_______________________________________________________________________________________________
ON X X X X 234.4 249.2 Normal BLAST
Field
OFF Data from
2/18/4
_______________________________________________________________________________________________
ON X X X X 201.6 253.1 Reverse BLAST
Field
OFF Data from
2/12/4
_______________________________________________________________________________________________
Data from 2/18/2004
BLAST at normal polarity = 6730 Amps
Transverse Holding Field = 500 Amps
Longitudinal Holding Field = 250 Amps
BRP setting = 170 Amps
Data was taken starting from the bottom of the table to the top
Holding Fields were cycled at each step
-- Shannon Krause Accelerator Operations Group Leader MIT Bates Linear Accelerator Center 21 Manning Rd. - P.O. Box 846 Middleton, MA 01949, USA617-253-9206 krause@mit.edu