No actual meeting today.
        Run plan for the weekend is to continue running ABS deuterium.  Keep an 
eye on the compression tube (orange trace) to be sure the ABS transitions 
are set correctly.  Keep pol.log and RUNLIST-D2.4 up to date.
        Genya hopes the nozzle will not freeze until Monday.  If it does freeze 
over the weekend, call him for instructions on how to regenerate it.  While 
it regenerates, we will run empty target.  Symptoms of nozzle freeze: ligit 
drops, line pressure (light blue line on the vacuum strip chart on 
dblast14) increases, C1 goes above 9.3 Torr, coincidence rate (green trace 
on BQM strip chart, green number on the lowest box in that EPICS screen) 
drops below 0.08 counts/Coulomb.
                                        Karen
This archive was generated by hypermail 2.1.2 : Mon Feb 24 2014 - 14:07:32 EST